LIGO Document P1900244-v12
- The precise calibration of the strain readout of the LIGO gravitational wave observatories is paramount to the accurate interpretation of gravitational wave events. This calibration is traditionally done by imparting a known force on one of the test masses of the interferometer via radiation pressure. Here we describe the implementation of an alternative calibration scheme: the Newtonian Calibrator. This system uses a rotor consisting of both quadrupole and hexapole mass distributions to apply a time-varying gravitational force on the test mass. The force produced by this rotor can be predicted to \( 1\% \) relative uncertainty and is well resolved in the readout of the interferometer. This system currently acts as a cross-check of the existing absolute calibration system.
Git document repository here: https://git.ligo.org/timesh.mistry/ncal-publication
Published 27 October 2021
Cite as: https://journals.aps.org/prd/abstract/10.1103/PhysRevD.104.082006
arXiv version:
https://arxiv.org/abs/2107.00141
DCC Version 3.4.3, contact
Document Database Administrators