LIGO Document G2100331-v1

G1 Calibrated Sensitivity Spectra Apr 11 2020 (Representative best of O3GK -- C02)

Document #:
LIGO-G2100331-v1
Document type:
G - Presentations (eg Graphics)
Other Versions:
Abstract:
Calibrated strain spectra for the GEO 600 detector from a representative best of O3GK Science run, taken on Apr 11, 2020, when the 1.4-1.4 binary system's luminosity range was 1 Mpc.

This data is based on offline recalibration (C02), which compensates for time-dependence in the detector response. Cleaning is carried out through the subtraction of dominant linear & bilinear noise contributions.

We expect the uncertainty in calibration at this time to be 10% in amplitude and 10 degrees in phase, valid between 40 and 6000 [Hz] (the uncertainty is larger and unquantified outside this band).

Keywords:
O3GK GEO600