PUNCH LIST - NONE. ALL GOOD
SIGNED - Calum Torrie signature on 2nd April 2014 closes this Acceptance Doc on behalf of chair and committee .
Punchlist items and their resolution will be tracked in T1400181. They will not be updated on this Acceptance Document filecard.
II. Documents for Review (against list from M1100282)
Top Assembly Name:
DOES NOT EXIST YET AdvLIGO SUS HAM5-I1, XYZ Local CS for HLTS (SR3)
QUESTION TO REVIEWERS - Will systems be creating layouts for I1? Or do we "accept" it without it?
ACCEPT WITHOUT. CIT. SYSTEMS WILL NOT BE CREATING I1 LAYOUT AT THIS POINT.
ICS Assembly #:
- ICS Assembly Load: https://ics-redux.ligo-la.caltech.edu/JIRA/browse/ASSY-D070447-005
Acceptance document:
This filecard serves as the acceptance document for 3rd IFO PR3.
For common material for all HLTS suspensions, see https://dcc.ligo.org/LIGO-E1201041
ICS inventory
ACTION - All to "find" an item and comment on success and structure
ACCEPTED. CLOSED.
S-numbers
Electronics for I1 have been inventoried, but Rack Layouts and/or individual S numbers have not yet been assigned to specific HSTSs.
QUESTION TO REVIEWERS - Is this OK? Should we dig up chassis test reports performed by CDS? I think chassis are being stored in Racks in the H2 Electronics Room; seems like we could treat them as though that was their "installed" location, and treat the inventory as LLO has with a DCC tree structure.
ACCEPT WITHOUT. CIT.
Test results:
- Calibrated, M1toM1 Transfer Functions (DAMPING ON/OFF):
https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=5289
- Calibrated, M2toM2 Transfer Functions (DAMPING ON/OFF):
Not applicable for 3rd IFO suspension.
- Calibrated, M3toM3 Transfer Functions (DAMPING ON/OFF):
Not applicable for 3rd IFO suspension.
- Calibrated OSEM Sensor ASDs, All Stages (DAMPING ON/OFF):
https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=5289
- B&K Hammer Cage Resonance Transfer Functions:
Not applicable for 3rd IFO suspension.
- Coil Driver Noise: https://alog.ligo-la.caltech.edu/aLOG/index.php?callRep=4751
- Coil Driver Response in all states: https://alog.ligo-la.caltech.edu/aLOG/index.php?callRep=4495
Noise measurements presented are for MC2, however, SR3 uses the same coil drivers and we would therefore expect identical performance.
PREVIOUSLY REVIEWED AND INCLUDED FOR COMPLETENESS.
- Violin Mode Frequencies and Q
Measured violin modes: https://alog.ligo-la.caltech.edu/aLOG/index.php?callRep=9418
- Summary of modeled violin modes: https://dcc.ligo.org/LIGO-T1300876
Measured violin mode Qs are used to estimate wire losses (structural and thermo-elastic) and hence expected thermal noise at lower frequencies. This falls outside of the scope of the fabrication acceptance review.
Channel/Signal List and Calibrations
For latest measurements of OSEM Sensor's Open Light Current, See https://dcc.ligo.org/LIGO-E1200343
Otherwise calibrations are assumed to be as shown in HLTS Control Design Description,
https://dcc.ligo.org/LIGO-T1000061 with a roughly 20% uncertainty
III. General notes and action items:
- The review team should review E1201041 in detail - unless the SUS Acceptance format changes and E1201041 is no longer the preferred method of presenting this info.