How to test:
- Use the VCO test rig D1100545.
- The excitation enable switch in principle will toggle the relay.
- However, a pull-up resistor of 1K needs to be added first between the open pin of the EXCEN_S switch and VCC (e.g., VCC pin of close-by LED)
- Turn both front panel power switches on (the two to the right).
- Toggle the relay and make sure the LEDs turn off.
- Check the readback by probing the voltage between pins 12/25 on the read DB25.
- It should be -6V for the most right switch, and -3V for the middle switch, and all combinations.