file://localhost
LIGO Production Large Actuator Test Report |
Actuator Serial #: L138
Operator Name: Gordon, Matt
Date: 4/12/2010 Time: 4:35 PM
Actuator Coil Resistance: 6.33 Ohms, PASS
Ambient Temperature: 73.3 F
Hi Pot Test Results: 1000 MOhms, PASS
X Travel Limit (inches): 0.537
Y Travel Limit (inches): 0.206
Z Travel Limit (inches): 0.509
|
X Position (in) |
Y Position (in) |
Z Position (in) |
Current (A) |
Force (lbs) |
Force Constant (lbs/A) |
PASS/FAIL |
Center |
0.000 |
0.000 |
0.000 |
2.077 |
19.953 |
9.606 |
PASS |
Position 1 |
0.250 |
0.080 |
0.217 |
2.475 |
20.077 |
8.111 |
PASS |
Position 2 |
0.250 |
-0.080 |
0.217 |
2.488 |
20.046 |
8.057 |
PASS |
Position 3 |
-0.250 |
-0.080 |
0.217 |
2.451 |
19.935 |
8.133 |
PASS |
Position 4 |
-0.250 |
0.080 |
0.217 |
2.368 |
19.829 |
8.373 |
PASS |
Position 5 |
-0.250 |
0.080 |
-0.217 |
2.439 |
20.061 |
8.224 |
PASS |
Position 6 |
-0.250 |
-0.080 |
-0.217 |
2.403 |
20.137 |
8.379 |
PASS |
Position 7 |
0.250 |
-0.080 |
-0.217 |
2.374 |
20.095 |
8.464 |
PASS |
Position 8 |
0.250 |
0.080 |
-0.217 |
2.446 |
20.100 |
8.219 |
PASS |
Center |
0.000 |
0.000 |
0.000 |
-2.181 |
-20.035 |
9.186 |
PASS |
Position 1 |
0.250 |
0.080 |
0.217 |
-2.378 |
-19.842 |
8.345 |
PASS |
Position 2 |
0.250 |
-0.080 |
0.217 |
-2.380 |
-19.890 |
8.357 |
PASS |
Position 3 |
-0.250 |
-0.080 |
0.217 |
-2.383 |
-19.963 |
8.378 |
PASS |
Position 4 |
-0.250 |
0.080 |
0.217 |
-2.389 |
-19.808 |
8.292 |
PASS |
Position 5 |
-0.250 |
0.080 |
-0.217 |
-2.483 |
-20.142 |
8.111 |
PASS |
Position 6 |
-0.250 |
-0.080 |
-0.217 |
-2.393 |
-19.937 |
8.332 |
PASS |
Position 7 |
0.250 |
-0.080 |
-0.217 |
-2.383 |
-20.250 |
8.498 |
PASS |
Position 8 |
0.250 |
0.080 |
-0.217 |
-2.436 |
-20.141 |
8.267 |
PASS |
QNA-TSG
1901 South Harbor City Blvd
Suite 700
Melbourne, FL 32901
Contact: Scott Stewart
scott.stewart@plansys.com
321-768-6500 x11
© 2009 QinetiQ North America, Technology Solutions Group |