file://localhost
LIGO Production Large Actuator Test Report |
Actuator Serial #: L041
Operator Name: Hartmann Donna
Date: 9/23/2009 Time: 10:05 AM
Actuator Coil Resistance: 6.37 Ohms, PASS
Ambient Temperature: 72.9 F
Hi Pot Test Results: 1000 MOhms, PASS
X Travel Limit (inches): 0.523
Y Travel Limit (inches): 0.207
Z Travel Limit (inches): 0.506
|
X Position (in) |
Y Position (in) |
Z Position (in) |
Current (A) |
Force (lbs) |
Force Constant (lbs/A) |
PASS/FAIL |
Center |
0.000 |
0.000 |
0.000 |
2.062 |
20.496 |
9.939 |
PASS |
Position 1 |
0.250 |
0.080 |
0.217 |
2.409 |
20.014 |
8.309 |
PASS |
Position 2 |
0.250 |
-0.080 |
0.217 |
2.389 |
20.035 |
8.386 |
PASS |
Position 3 |
-0.250 |
-0.080 |
0.217 |
2.511 |
20.050 |
7.985 |
PASS |
Position 4 |
-0.250 |
0.080 |
0.217 |
2.479 |
20.348 |
8.209 |
PASS |
Position 5 |
-0.250 |
0.080 |
-0.217 |
2.262 |
20.263 |
8.958 |
PASS |
Position 6 |
-0.250 |
-0.080 |
-0.217 |
2.328 |
20.281 |
8.713 |
PASS |
Position 7 |
0.250 |
-0.080 |
-0.217 |
2.327 |
20.256 |
8.704 |
PASS |
Position 8 |
0.250 |
0.080 |
-0.217 |
2.371 |
20.256 |
8.545 |
PASS |
Center |
0.000 |
0.000 |
0.000 |
-2.146 |
-20.173 |
9.399 |
PASS |
Position 1 |
0.250 |
0.080 |
0.217 |
-2.386 |
-20.300 |
8.507 |
PASS |
Position 2 |
0.250 |
-0.080 |
0.217 |
-2.385 |
-20.262 |
8.495 |
PASS |
Position 3 |
-0.250 |
-0.080 |
0.217 |
-2.337 |
-20.026 |
8.568 |
PASS |
Position 4 |
-0.250 |
0.080 |
0.217 |
-2.381 |
-20.276 |
8.516 |
PASS |
Position 5 |
-0.250 |
0.080 |
-0.217 |
-2.443 |
-20.300 |
8.308 |
PASS |
Position 6 |
-0.250 |
-0.080 |
-0.217 |
-2.237 |
-20.301 |
9.076 |
PASS |
Position 7 |
0.250 |
-0.080 |
-0.217 |
-2.465 |
-20.121 |
8.161 |
PASS |
Position 8 |
0.250 |
0.080 |
-0.217 |
-2.433 |
-20.353 |
8.364 |
PASS |
QNA-TSG
1901 South Harbor City Blvd
Suite 700
Melbourne, FL 32901
Contact: Scott Stewart
scott.stewart@plansys.com
321-768-6500 x11
© 2009 QinetiQ North America, Technology Solutions Group |