LIGO Production Large Actuator Test Report |
Actuator Serial #: L016
Operator Name: Smith, Lane
Date: 8/12/2009 Time: 1:06 PM
Actuator Coil Resistance: 6.42 Ohms, PASS
Ambient Temperature: 69.3 F
Hi Pot Test Results: 1000 MOhms, PASS
X Travel Limit (inches): 0.521
Y Travel Limit (inches): 0.205
Z Travel Limit (inches): 0.506
|
X Position (in) |
Y Position (in) |
Z Position (in) |
Current (A) |
Force (lbs) |
Force Constant (lbs/A) |
PASS/FAIL |
Center |
0.000 |
0.000 |
0.000 |
2.106 |
20.286 |
9.633 |
PASS |
Position 1 |
0.250 |
0.075 |
0.215 |
2.505 |
20.056 |
8.005 |
PASS |
Position 2 |
0.250 |
-0.075 |
0.215 |
2.333 |
19.892 |
8.528 |
PASS |
Position 3 |
-0.250 |
-0.075 |
0.215 |
2.347 |
19.934 |
8.492 |
PASS |
Position 4 |
-0.250 |
0.075 |
0.215 |
2.451 |
20.012 |
8.163 |
PASS |
Position 5 |
-0.250 |
0.075 |
-0.215 |
2.335 |
20.200 |
8.650 |
PASS |
Position 6 |
-0.250 |
-0.075 |
-0.215 |
2.336 |
19.998 |
8.562 |
PASS |
Position 7 |
0.250 |
-0.075 |
-0.215 |
2.358 |
20.021 |
8.492 |
PASS |
Position 8 |
0.250 |
0.075 |
-0.215 |
2.178 |
19.824 |
9.104 |
PASS |
Center |
0.000 |
0.000 |
0.000 |
-2.102 |
-20.369 |
9.691 |
PASS |
Position 1 |
0.250 |
0.075 |
0.215 |
-2.310 |
-19.999 |
8.658 |
PASS |
Position 2 |
0.250 |
-0.075 |
0.215 |
-2.317 |
-20.224 |
8.730 |
PASS |
Position 3 |
-0.250 |
-0.075 |
0.215 |
-2.303 |
-20.060 |
8.712 |
PASS |
Position 4 |
-0.250 |
0.075 |
0.215 |
-2.295 |
-19.924 |
8.682 |
PASS |
Position 5 |
-0.250 |
0.075 |
-0.215 |
-2.422 |
-20.169 |
8.328 |
PASS |
Position 6 |
-0.250 |
-0.075 |
-0.215 |
-2.401 |
-19.855 |
8.270 |
PASS |
Position 7 |
0.250 |
-0.075 |
-0.215 |
-2.413 |
-20.021 |
8.298 |
PASS |
Position 8 |
0.250 |
0.075 |
-0.215 |
-2.354 |
-20.140 |
8.556 |
PASS |