LIGO Production Large Actuator Test Report |
Actuator Serial #: L020
Operator Name: Smith, Lane
Date: 8/12/2009 Time: 10:32 AM
Actuator Coil Resistance: 6.34 Ohms, PASS
Ambient Temperature: 70.6 F
Hi Pot Test Results: 1000 MOhms, PASS
X Travel Limit (inches): 0.521
Y Travel Limit (inches): 0.204
Z Travel Limit (inches): 0.508
|
X Position (in) |
Y Position (in) |
Z Position (in) |
Current (A) |
Force (lbs) |
Force Constant (lbs/A) |
PASS/FAIL |
Center |
0.000 |
0.000 |
0.000 |
2.155 |
20.235 |
9.391 |
PASS |
Position 1 |
0.250 |
0.075 |
0.215 |
2.496 |
19.844 |
7.951 |
PASS |
Position 2 |
0.250 |
-0.075 |
0.215 |
2.423 |
19.931 |
8.226 |
PASS |
Position 3 |
-0.250 |
-0.075 |
0.215 |
2.372 |
19.943 |
8.406 |
PASS |
Position 4 |
-0.250 |
0.075 |
0.215 |
2.506 |
19.897 |
7.940 |
PASS |
Position 5 |
-0.250 |
0.075 |
-0.215 |
2.455 |
20.077 |
8.178 |
PASS |
Position 6 |
-0.250 |
-0.075 |
-0.215 |
2.407 |
20.203 |
8.395 |
PASS |
Position 7 |
0.250 |
-0.075 |
-0.215 |
2.394 |
20.239 |
8.456 |
PASS |
Position 8 |
0.250 |
0.075 |
-0.215 |
2.301 |
20.110 |
8.741 |
PASS |
Center |
0.000 |
0.000 |
0.000 |
-2.031 |
-19.845 |
9.769 |
PASS |
Position 1 |
0.250 |
0.075 |
0.215 |
-2.302 |
-20.095 |
8.730 |
PASS |
Position 2 |
0.250 |
-0.075 |
0.215 |
-2.279 |
-20.138 |
8.838 |
PASS |
Position 3 |
-0.250 |
-0.075 |
0.215 |
-2.277 |
-19.835 |
8.711 |
PASS |
Position 4 |
-0.250 |
0.075 |
0.215 |
-2.315 |
-20.162 |
8.709 |
PASS |
Position 5 |
-0.250 |
0.075 |
-0.215 |
-2.541 |
-19.917 |
7.837 |
PASS |
Position 6 |
-0.250 |
-0.075 |
-0.215 |
-2.467 |
-20.291 |
8.224 |
PASS |
Position 7 |
0.250 |
-0.075 |
-0.215 |
-2.492 |
-19.975 |
8.017 |
PASS |
Position 8 |
0.250 |
0.075 |
-0.215 |
-2.358 |
-19.958 |
8.464 |
PASS |